Troubleshooting the Coolifting Device - Coolifting Gun is Clogged - Trigger is Stuck

Troubleshooting the Coolifting Device - Coolifting Gun is Clogged - Trigger is Stuck

This article outlines the procedure set by the Coolifting Manufacturer to unclog a Coolifting Gun.
This procedure will create negative pressure in the device and force CO2 to flow through the serum puncture needles.
  1. While performing a Coolifting procedure, remove the serum vial and block the CO2 gas output (device tip) by facing the tip against a flat surface. Keep pressing the trigger for a few seconds.
  2. You will see any product left in the serum lines flow out of the serum puncture needles.
  3. Replace the serum vial and continue the treatment normally. The serum distribution will function normally now.   
 
This video visually demonstrates the procedures explained above:





Here is a video demonstrating how to unstick the trigger:





Should you have any questions please contact ProMed Solutions at (562) 326-5720 or e-mail us at support@promedsi.com.

As always, thank you for your business!

    • Related Articles

    • CooLifting Cleaning Kit - A kit to be used to prevent your CooLifting device from getting clogged

      When placing a new CooLifting order online (shop.promedsi.com), you can request to add a CooLifting Cleaning Kit to your order free of charge. Just place a note at checkout and we will include a CooLifting Cleaning Kit with your order. The CooLifting ...
    • Coolifting 2.0 User Manual

      Please download the attachment found in this article to view the Coolifting 2.0 User Manual as a PFD. Should you have any questions please contact ProMed Solutions at (562) 326-5720 or e-mail us at support@promedsi.com. As always, thank you for your ...
    • Coolifting- How to locate your serial number

      This article outlines to procedure to locate your Coolifting device's serial number. Coolifting Gun: 1. Locate the CO2 insertion point. 2. The serial number is printed on a label and is located on the left, inside of the CO2 insertion point on the ...
    • Coolifting Informed Consent Forms

      Attached are two different Informed Consent forms.
    • Coolifting 2.0 Marketing Photos - Photos Gun, Base, and Product

      Feel free to use any of the marketing stock images bellow. Right click on the image desired and save as an image locally to your computer. Should you have any questions please contact ProMed Solutions at (562) 326-5720 or e-mail us ...